Technical specifications |
||||
Pressure specifications: 0.02% of full scale (730RG), 0.05% of full scale (730G) |
||||
Current specifications: 0.01% of reading |
||||
Model |
Range |
Resolution |
Accuracy1 |
Burst pressure |
730G01 |
-10 to +10 inH2O -20 to 20 mbar -2.49 to 2.49 kPa |
0.001 inH2O 0.001 mbar 0.0001 kPa |
Positive pressure, ±0.2% of range2 |
3 psi, 210 mbar 20.6 kPa |
730G02 |
-1 to +1 psi -70 to 70 mbar -6.89 to 6.89 kPa |
0.0001 psi 0.001 mbar 0.0001 kPa |
5 psi, 350 mbar 34.4 kPa |
|
730G04 |
-14 to 15 psi -0.97 to 1 bar -97 to 103 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
Positive pressure, ±0.05% of range Vacuum, ±0.1% of range |
60 psi 4 bar 413.6 kPa |
730G05 |
-14 to 30 psi -0.97 to 2 bar -97 to 206 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
120 psi 8 bar 827.3 kPa |
|
730G06 |
-12 to 100 psi -0.83 to 6.9 bar -83 to 690 kPa |
0.01 psi, 0.0001 bar |
400 psi 26 bar 2.757 MPa |
|
730G27 |
-12 to 300 psi -0.83 to 20 bar -83 to 2 MPa |
0.01 psi 0.001 bar 0.1 kPa |
1200 psi 80 bar 8.273 MPa |
|
730G07 |
-12 to 500 psi -0.83 to 34 bar -83 to 3.4 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 13.789 MPa |
|
730G08 |
-14 to 1000 psi, -0.97 to 69 bar -97 to 6900 kPa |
0.01 psi 0.001 bar 0.1 kPa |
4000 psi 266 bar 27.579 MPa |
|
730G10 |
-14 to 2000 psi, -0.97 to 140 bar -97 to 14 MPa |
0.1 psi 0.01 bar 1 kPa |
8000 psi 550 bar 55.158 MPa |
|
730G29 |
-14 to 3000 psi, -0.97 to 200 bar -97 to 20 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730G30 |
-14 to 5000 psi, -0.97 to 340 bar -97 to 34 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730G31 |
-14 to 10000 psi, -0.97 to 690 bar -97 to 69 MPa |
1 psi 0.01 bar 1 kPa |
15000 psi 1035 bar 103.421 MPa |
|
730GA4 |
0 to 15 psia, 0 to 1 bar 0 to 100 kPa absolute pressure |
0.001 psi 0.0001 bar 0.01 kPa |
±0.05% of range4 |
60 psi 4 bar 413.6 kPa |
730GA5 |
0 to 30 psia 0 to 2 bar 0 to 200 kPa absolute pressure |
0.001 psi 0.0001 bar 0.01 kPa |
120 psi 8 bar 827.3 kPa |
|
730GA6 |
0 to 100 psia, 0 to 6.9 bar 0 to 690 kPa absolute pressure |
0.01 psi 0.0001 bar 0.01 kPa |
400 psi 27 bar 2.757 MPa |
|
730GA27 |
0 to 300 psia 0 to 20 bar 0 to 2 MPa absolute |
0.01 psi 0.001 bar .01 kPa |
1200 psi 80 bar 8.273 MPa |
|
730RG04 |
-14 to 15 psi, -0.97 to 1 bar -97 to 100 kPa |
0.001 psi 0.0001 bar 0.0001 kPa |
Positive pressure, ±0.02 % of range3 Vacuum, ±0.05 % of range3 |
45 psi 3 bar 310.2 kPa |
730RG05 |
-14 to 30 psi, -0.97 to 2 bar -97 to 200 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
90 psi 6 bar 620.5 kPa |
|
730RG06 |
-12 to 100 psi, -0.83 to 6.9 bar -83 to 690 kPa |
0.01 psi 0.0001 bar 0.01 kPa |
1000 psi 69 bar 6.894 MPa |
|
730RG27 |
–12 psi to 300 psi -0.83 to 20 bar –80 kPa to 2 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 14 MPa |
|
730RG07 |
–12 psi to 500 psi -0.83 to 34 bar –80 kPa to 3.5 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 13.789 MPa |
|
730RG08 |
-14 to 1000 psi, -0.97 to 69 bar -97 to 6900 kPa |
0.1 psi 0.001 bar 0.1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG29 |
-14 to 3000 psi -0.97 to 200 bar -97 to 20 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG30 |
-14 to 5000 psi, -0.97 to 340 bar -97 to 34 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG31 |
-14 to 10000 psi, -0.97 to 690 bar -97 to 69 MPa |
0.1 psi 0.01 bar 1 kPa |
15000 psi 1035 bar 103.421 MPa |
|
1 Accuracy specifications above have considered the effects of linearity, repeatability, resolution, reference accuracy, 1-year stability, and temperature effect. 2 For 730G01/730G02, accuracy is ±0.3% of full scale @ 0 °C to 15 °C and 35 °C to 50 °C. 3 For all 730RG model, add 0.003% FS/ °C @ 0 °C to 18 °C and 28 °C to 50 °C。 4 For 730GA4, add 0.006% FS/ °C @ -10 °C to 15 °C and 35 °C to 55 °C; for all other 730GA model, add 0.003% FS/ °C @ -10 °C to 15 °C and 35 °C to 55 °C. |
||||
Media Compatibility 730G01, 730G02, 730G04, 730G05, 730RG04, 730RG05: any clean dry non-corrosive gas Above 100 psi: any liquids or gases compatible with 316 stainless steel |
Electrical specifications (All specifications are valid to 110% of range) |
|
Current |
0 mA to 24 mA |
Voltage |
0 V DC to 30 V DC |
Current range |
1 μA |
Voltage range |
1 mV |
Accuracy |
0.01% ± 2 LSD all ranges @ 23 °C ±5 °C |
Temperature compensation |
From 20% FS/ °C @ -10 °C to +18 °C and 28 °C to 50 °C |
Loop Compliance Voltage |
24 V DC @ 20 mA |
Temperature measurement accuracy inside the product |
±0.5 ℃ |
Drive Capability |
1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |